MAUDE Adverse Event Report: COVIDIEN MFG DC BOULDER LIGASURE; ELECTROSURGICAL, CUTTING & COAGULATION & ACCESSORIES

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

COVIDIEN MFG DC BOULDER LIGASURE; ELECTROSURGICAL, CUTTING & COAGULATION & ACCESSORIES Back to Search Results
Model Number LF1544
Medical Device Problem Code Separation Problem (4043)
Health Effect - Clinical Code No Consequences Or Impact To Patient (2199)
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Evaluation summary: one device was received for evaluation.The returned product did not meet specification as received.Visual inspection found the cord plug was cut off and not returned.The reported condition could not be confirmed.Investigation found that the cord has been cut off and the cord plug was not returned.Without the cord plug, a detail investigation could not be performed for the reported complaint.The investigation could not determine the root cause of the customer's report.If information is provided in the future, a supplemental report will be issued.
 
Event or Problem Description
According to the reporter, during a procedure, there was an excessive sticking on tissue on the device.There was no patient injury.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
LIGASURE
Common Device Name
ELECTROSURGICAL, CUTTING & COAGULATION & ACCESSORIES
Manufacturer (Section D)
COVIDIEN MFG DC BOULDER
5920 longbow dr
boulder CO 80301 3299
MDR Report Key9104766
Report Number1717344-2019-01200
Device Sequence Number1912739
Product Code GEI
UDI-Device Identifier10884521124875
UDI-Public10884521124875
Combination Product (Y/N)N
Initial Reporter CountryAS
PMA/510(K) Number
K092879
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,foreig
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date (Section B) 09/23/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberLF1544
Device Catalogue NumberLF1544
Device Lot Number51120086X
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer09/04/2019
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 08/27/2019
Initial Report FDA Received Date09/23/2019
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-