COVIDIEN FOLEY TRAY SILICON DRAIN BAG 16FR; CATHETER, RETENTION TYPE, BALLOON Back to Search Results
Model Number 6146LL
Medical Device Problem Code Disconnection (1171)
Health Effect - Clinical Code No Code Available (3191)
Date of Event 08/28/2019
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
The incident sample has been requested but to date has not been received for evaluation.If the sample is received, or if additional information pertinent to the incident is obtained a follow-up report will be submitted.As part of our manufacturing process, all device history records are reviewed and approved by quality, prior to release of product.
 
Event or Problem Description
The customer reported that upon removal of catheter, about 3/4 inch of the end of the catheter was missing.Additional information received from the customer stated that the catheter was in place less than 24 hours and removed after delivery.It was used on a female patient.A flex cystoscopy procedure was performed to look for the missing end of the catheter.No missing end of the catheter was found inside the patient.The patient is currently doing well with no issues.
 
Additional Manufacturer Narrative
A device history record review could not be performed because a valid cardinal health lot number was not received with the complaint.As part of our manufacturing process, all device history records are reviewed and approved by quality, prior to the release of product.One decontaminated sample was received at the manufacturing site for the investigation.The sample received was damaged, as a result testing could not be completed to confirm the reported issue.A review of the manufacturing process of the drainage bag has concluded that the reported condition could not have been generated in the production line at cardinal health.The catheter is a purchased component provided by an external supplier.The catheter arrives inside of a sleeve prior to assembling it with the pvc tubing.The operator does not remove the sleeve during this process.The same process is followed in the workstation (sub-assembly area) where the catheter is assembled with the adapter, again the sleeve is not removed.Therefore, the reported condition could not be generated during the manufacturing process of the drainage bag.A complaint notification was sent to the external supplier along with the sample to initiate the investigation of the root cause of the reported condition and request corrective actions.This complaint will be used for qa tracking and trending purposes.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
FOLEY TRAY SILICON DRAIN BAG 16FR
Common Device Name
CATHETER, RETENTION TYPE, BALLOON
Manufacturer (Section D)
COVIDIEN
calle 9 sur no. 125 cuidad ind
tijuana 22500
MX  22500
MDR Report Key9105791
Report Number9612030-2019-02273
Device Sequence Number13991216
Product Code EZL
UDI-Device Identifier10884521015500
UDI-Public10884521015500
Combination Product (Y/N)N
Initial Reporter StateIA
Initial Reporter CountryUS
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,user f
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 02/28/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device No Information
Device Model Number6146LL
Device Catalogue Number6146LL
Device Lot Number5114268
Was Device Available for Evaluation? Yes
Initial Date Received by Manufacturer 09/06/2019
Supplement Date Received by Manufacturer09/06/2019
Initial Report FDA Received Date09/23/2019
Supplement Report FDA Received Date02/28/2020
Patient Sequence Number1
Outcome Attributed to Adverse Event Other;
-
-