| Catalog Number |
8559H |
| Medical Device Problem Code |
Break (1069)
|
| Health Effect - Clinical Code |
No Consequences Or Impact To Patient (2199)
|
| Date of Event |
08/27/2019
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
(b)(4).Attempts are being made to retrieve the device.To date the device has not been returned.If the device or further details are received at a later date a supplemental medwatch will be sent.
|
| |
|
Event or Problem Description
|
|
It was reported that the patient underwent cabg procedure on (b)(6) 2019 and suture was used.During a coronary artery bypass graft, the needle was broken at the swage part during suturing the vein.The needle piece was removed without losing sight.No pieces were left inside the patient.There were no adverse patient consequences reported.
|
| |
|
Additional Manufacturer Narrative
|
|
Product complaint # (b)(4).A manufacturing record evaluation was performed for the finished device batch mmh523, and no non-conformances were identified.Device evaluation summary: a suture needle was submitted for evaluation.A fracture was observed at the suture attachment of the needle.The needle was received in two pieces, only one of the mating fracture surfaces was examined for this evaluation.A scanning electron microscope (sem) was used to examine the fracture surfaces and surrounding area of the needle.The fracture surfaces were examined in multiple locations in order to determine the fracture mode.The evaluation revealed the fracture was composed of microvoid coalescence, which is evidence of a ductile overload failure.This was a ductile fracture.The manufacturing records were reviewed, and the manufacturing/packaging criteria were met prior to the release of this batch.The evidence of this examination indicates that the breakage occurred at the attachment area of the needle during use due to tensile overload.There is no evidence of any material flaw or defect that would cause premature failure.In order to avoid this kind of damage grasp the needle in an area one-third (1/3) to one-half (1/2) of the distance from the attachment end to the point.
|
| |
|
Search Alerts/Recalls
|