• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: US SURGICAL PUERTO RICO EEA; STAPLE, IMPLANTABLE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

US SURGICAL PUERTO RICO EEA; STAPLE, IMPLANTABLE Back to Search Results
Model Number EEAXL2835
Medical Device Problem Code Gas/Air Leak (2946)
Health Effect - Clinical Codes Tissue Damage (2104); No Code Available (3191)
Date of Event 08/28/2019
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
(b)(4).If information is provided in the future, a supplemental report will be issued.
 
Event or Problem Description
According to the reporter, during open, low anterior resection procedure, while an end to end anastomosis, the surgeon observed bubbles of air from the staple line of the initial anastamosis but the staples did form and the tissue was correctly cut.Over sewed was done, then a second device was opened and another anastamosis was performed where there did not appear to be a distal donut.Surgical time was extended and there was unanticipated tissue loss and tissue damage reported.
 
Additional Manufacturer Narrative
Evaluation summary: post market vigilance (pmv) led an evaluation of two devices.The visual inspection of both staple guides noted the instruments were fully applied.Microscopic examination noted nicks in both knife blades.The anvils of both instruments were observed to be tilted and separated from their respective instrument.For instrument #1, a shallow knife impression was observed along the cutline impression, with the cut ring found to be partially severed.For instrument #2, a deep knife impression was observed along the cutline impression, with the cut ring found to be completely severed.Functionally, both devices were applied over the appropriate test media producing acceptable results.The knives cut the test media cleanly and completely, and the pushers advanced, despite the noted knife blade damage.Records from each manufacturing lot are thoroughly reviewed to ensure that products are released meeting all quality release specifications at the time of manufacture.Replication of the reported condition may occur if the instrument handle compression is not completed or if the instrument is applied against an excessive amount of tissue during the clinical application.Also, the instructions for use of this product states: when firing the stapler, make sure to fully squeeze the instrument handle until the metal underside of the handle contacts the stapler body to the fullest extent.Partial or incomplete handle squeezes may result in unacceptable staple formation and/or incomplete knife cut.The root cause of the observed damage was found to be due to the device not being used as intended which caused or contributed to the reported condition.No further actions have been deemed necessary at this time.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
EEA
Common Device Name
STAPLE, IMPLANTABLE
Manufacturer (Section D)
US SURGICAL PUERTO RICO
201 sabanetas industrial park
ponce PR 00716 4401
MDR Report Key9107760
Report Number2647580-2019-04728
Device Sequence Number1924350
Product Code GDW
UDI-Device Identifier20884523006145
UDI-Public20884523006145
Combination Product (Y/N)N
Initial Reporter StateGA
Initial Reporter CountryUS
PMA/510(K) Number
K062850
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2019
Reporter Type Manufacturer
Report Source company representative,health
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 12/17/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date02/29/2024
Device Model NumberEEAXL2835
Device Catalogue NumberEEAXL2835
Device Lot NumberP9C1489KY
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer09/11/2019
Initial Date Received by Manufacturer 08/28/2019
Supplement Date Received by Manufacturer12/05/2019
Initial Report FDA Received Date09/23/2019
Supplement Report FDA Received Date12/17/2019
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
-
-