Lot Number 5208243
Medical Device Problem Code Detachment of Device or Device Component (2907)
Health Effect - Clinical Code Elevated ketones/Diabetic Ketoacidosis (2364)
Type of Reportable Event Malfunction
Event or Problem Description
Unomedical reference number (b)(4).Event occurred in (b)(6).It was reported that patient's mother noticed an infusion set's tubing detachment from the site location at quick release which caused high blood glucose level.This issue occurred five times and patient's mother saved three used sets that had the detachment.The site location was at patient's arm and the pump was located at the same side to the site at the waist.The infusion had been in use for less than 24 hours.Reportedly, the infusions were not stored, or used, in a place where they might have been exposed to extreme temperatures and humidity.There was no stress or pull on the tubing and the pump was not dropped with the set connected to patient's body.No further information available.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
QUICK-SET® PARADIGM®
Common Device Name
QUICK-SET 60/6 PCC
MDR Report Key9108164
Report Number3003442380-2019-01817
Device Sequence Number14246394
Product Code FPA
UDI-Device Identifier05705244006317
UDI-Public05705244006317
Combination Product (Y/N)N
PMA/510(K) Number
K160648
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source distributor
Initial Reporter Occupation Other
Type of Report Initial
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device No Information
Device Expiration Date12/01/2020
Device Lot Number5208243
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 09/12/2019
Initial Report FDA Received Date09/24/2019
Patient Sequence Number1
-
-