| Model Number |
MMT-522NAS |
| Medical Device Problem Code |
No Display/Image (1183)
|
| Health Effect - Clinical Code |
No Consequences Or Impact To Patient (2199)
|
| Date of Event |
09/16/2019
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
(b)(4).Currently it is unknown whether or not the device may have caused or contributed to the event as no product has been returned.The device will be returned for analysis and further information will follow once the analysis has been completed.No conclusion can be drawn at this time.
|
| |
|
Event or Problem Description
|
|
Customer reported via phone call that the insulin pump had blank display.The customer¿s blood glucose level was 135 mg/dl.The customer reported that the display was not blank less than 30 seconds and did not return.The customer was assisted with troubleshooting and customer reports display was blank currently.The customer removes the battery and stated the insert battery alarm did not display on the insulin pump screen.Customer states the contact on the battery cap was neither missing nor damage.Customer states battery compartment was neither damaged nor corroded.Customer states the spring was neither damaged nor corroded.Customer states the display did not return after insulin pump restart.The customer was advised to discontinue use of the insulin pump and to revert to the back-up plan.The insulin pump will be returned for analysis.
|
| |
|
Additional Manufacturer Narrative
|
|
Device received with blank display due to corroded battery tube spring contact and corroded battery tube.Unable to perform all functional testing including the operating currents, a21 error test, rewind, basic occlusion, occlusion, prime/a33 and excessive no delivery test due to blank display.
|
| |
|
Search Alerts/Recalls
|