MAUDE Adverse Event Report: STRYKER ORTHOPAEDICS-MAHWAH UNKNOWN 6MM TIBIA OFFSET; PROSTHESIS, KNEE, PATELLO/FEMOROTIBIAL, SEMI-CONSTRAINED, UNCEMENTED, POROUS, CO

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

STRYKER ORTHOPAEDICS-MAHWAH UNKNOWN 6MM TIBIA OFFSET; PROSTHESIS, KNEE, PATELLO/FEMOROTIBIAL, SEMI-CONSTRAINED, UNCEMENTED, POROUS, CO Back to Search Results
Catalog Number UNK_JR
Device Problem Osseointegration Problem (3003)
Patient Problems Injury (2348); Inadequate Osseointegration (2646)
Event Date 08/27/2019
Event Type  Injury  
Manufacturer Narrative
An event regarding loosening involving tibial offset was reported.The event was not confirmed.Method & results: device evaluation and results: not performed as product was not returned.Clinician review: a review of the provided medical records by a clinical consultant stated the following comment: deemed insufficient for medical review [.] ¿ cannot confirm event, need additional information; primary and revision operative reports, clinical and past medical history, additional serial dated x-rays and examination of explanted components.Device history review: could not be performed as lot code information was not provided.Complaint history review: could not be performed as lot code information was not provided conclusion: the exact cause of the event could not be determined because insufficient information was provided.Additional information including operative reports, progress notes, x-rays and return of the device are needed to fully investigate the event.If further information becomes available or the product is returned, this investigation will be re-opened.Device not returned.
 
Event Description
It was reported that the patient's right knee was revised due to progressive radiolucency around the tibial and femoral stems.A knee construct with tibial baseplate and stem, femoral component and stem, and a cs insert were revised to a ts construct.Rep provided a webops report, an excerpt from an operative report, explant pictures and x-rays from (b)(6) 2019 and reported that no further information will be available.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
UNKNOWN 6MM TIBIA OFFSET
Type of Device
PROSTHESIS, KNEE, PATELLO/FEMOROTIBIAL, SEMI-CONSTRAINED, UNCEMENTED, POROUS, CO
Manufacturer (Section D)
STRYKER ORTHOPAEDICS-MAHWAH
325 corporate drive
mahwah NJ 07430
Manufacturer (Section G)
STRYKER ORTHOPAEDICS-MAHWAH
325 corporate drive
mahwah NJ 07430
Manufacturer Contact
collin neitzel
325 corporate drive
mahwah, NJ 07430
2018315000
MDR Report Key9109425
MDR Text Key164804789
Report Number0002249697-2019-03317
Device Sequence Number1
Product Code MBH
Combination Product (y/n)N
Reporter Country CodeUS
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type health professional,other
Reporter Occupation Physician
Type of Report Initial
Report Date 09/24/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received09/24/2019
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Health Professional
Device Catalogue NumberUNK_JR
Device Lot NumberUNKNOWN
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Date Manufacturer Received08/27/2019
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Hospitalization; Required Intervention;
Patient Age61 YR
-
-