Catalog Number 550514
Medical Device Problem Code Adverse Event Without Identified Device or Use Problem (2993)
Health Effect - Clinical Codes Foreign Body Reaction (1868); Hemorrhage/Blood Loss/Bleeding (1888); Hypersensitivity/Allergic reaction (1907); Inflammation (1932); Pain (1994); Tissue Damage (2104); Weakness (2145); Discomfort (2330); Limited Mobility Of The Implanted Joint (2671); No Code Available (3191)
Date of Event 08/23/2017
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Product complaint # (b)(4).Initial reporter occupation: lawyer.(b)(4).If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Event or Problem Description
Pinnacle claim submission form and medical records received.After review of medical records, the patient was revised to address dislocation, clicking, discomfort, bilateral lower extremity foot drop, pain, elevated blood metal ions and metal on metal reaction.Patient had stated that she felt that her hip has subluxed at times.Range of motion was noted to be limited.There was tenderness and weakness as well.Intraoperative findings include metallosis with "dish soap water fluid", a significantly abducted acetabular component and gross trunnionosis on the femoral head.During the revision surgery, patient had acute blood loss anemia secondary to the surgery.Two packed red blood cells were given.Doi: (b)(6) 2008.Dor: (b)(6) 2017, (right hip).
 
Additional Manufacturer Narrative
Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.Product complaint # (b)(4).No device associated with this report was received for examination.A worldwide complaint database search found no other related reported incidents against the provided product code/lot number combination since release for distribution.The information received will be retained for potential series investigations if triggered by trend analysis, post market surveillance, or other events within the quality system.Device history lot : null.Device history batch : null.Device history review : null.If information is obtained that was not available for the initial medwatch, a follow-up medwatch, a follow-up medwatch will be filed as appropriate.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Common Device Name
S-ROM HIP SYSTEM : HIP FEMORAL SLEEVE
Manufacturer (Section D)
DEPUY ORTHOPAEDICS INC US
700 orthopaedic drive
warsaw IN 46581 0988
MDR Report Key9110285
Report Number1818910-2019-105938
Device Sequence Number3041806
Product Code LPH
UDI-Device Identifier10603295175247
UDI-Public10603295175247
Combination Product (Y/N)N
Initial Reporter StateTX
Initial Reporter CountryUS
PMA/510(K) Number
K934412
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2008
Device Explanted Year2017
Reporter Type Manufacturer
Report Source other
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 09/03/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Catalogue Number550514
Device Lot Number2664921
Was Device Available for Evaluation? No
Initial Date Received by Manufacturer 09/03/2019
Supplement Date Received by Manufacturer10/22/2019
Initial Report FDA Received Date09/24/2019
Supplement Report FDA Received Date10/30/2019
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient Age48 YR
Patient Weight82
-
-