Catalog Number 284002
Device Problem Suction Failure (4039)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 08/23/2018
Event Type  malfunction  
Manufacturer Narrative
Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.Udi: (b)(4).Investigation summary: the device was received and evaluated at the service center.The complaint was not confirmed.The following information was derived from the service report: per service manual operational and diagnostic analysis does not confirm the reported issue (pinch valve on the suction side of the device did not have a sufficient seal on the tubing).Per sb, performed software upgrade to 3.11.Also replaced damage console cover.The testing of the unit was completed per the service manual.The unit passed all functional tests and is fully operational.Since the issue reported by the customer was not reproduced, no root cause is avialable for the reported failure.The console cover was most likely caused by user mishandling.At this point, no corrective action is required and no further action is warranted.However, depuy synthes mitek will continue to track any related complaints within this device family as a means of monitoring the extent with which this complaint is observed in the field.This report is being filed as required under mitek's corrective and preventative actions (capa) to file usa fda mdr missed malfunctions.
 
Event Description
It was reported by the sales rep via phone that during a shoulder repair procedure the customer's fms vue fluid management system pinch valve on the suction side of the device did not have a sufficient seal on the tubing.The sales rep stated the case was completed by holding this device to create better suction with no patient harm or delays.The sales rep was not present for the case and could not provide any additional information.There was no delay in the surgical procedure.There was a spare device available for use to complete the surgery.There was patient involvement.There were no reports of injuries, medical intervention or prolonged hospitalization.All available information has been disclosed.If additional information should become available, a supplemental medwatch will be submitted accordingly.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
FMS VUE PUMP
Type of Device
DISTENSION UNIT, FLUID, ARTHROSCOPIC
Manufacturer (Section D)
DEPUY MITEK LLC US
325 paramount drive
raynham MA 02767
Manufacturer (Section G)
DEPUY MITEK LLC US
325 paramount drive
raynham MA 02767
Manufacturer Contact
kara ditty-bovard
325 paramount drive
raynham, MA 02767
6013142063
MDR Report Key9110765
MDR Text Key207434124
Report Number1221934-2019-58473
Device Sequence Number1
Product Code HRX
UDI-Device Identifier10886705020591
UDI-Public10886705020591
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K130169
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,other
Reporter Occupation Other
Type of Report Initial
Report Date 08/23/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue Number284002
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer09/19/2018
Initial Date Manufacturer Received 08/23/2018
Initial Date FDA Received09/24/2019
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Unknown
Patient Sequence Number1
-
-