| Model Number |
775759 |
| Medical Device Problem Code |
Leak/Splash (1354)
|
| Health Effect - Clinical Code |
No Known Impact Or Consequence To Patient (2692)
|
| Date of Event |
04/29/2019
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The device history record review of the reported lot number shows evidence that the product was released according to all established procedures and qa documentation.Samples were received in its original package.Samples were visual and functional inspected.During the functional inspection a leak was detected between the connection of the tubing line and the cross-spike connector.The reported condition will be confirmed as related to a manufacturing / production process.The root cause and corrective / preventative action (capa) plan will be documented through a formal investigation and documented within the capa.
|
| |
|
Event or Problem Description
|
|
The customer reports the feeding tubes are leaking at spike connection, occurring with multiple sets.The leak is at the spike proximal to bag.
|
| |
|