• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ABBOTT VASCULAR STEERABLE GUIDE CATHETER; VALVE REPAIR

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ABBOTT VASCULAR STEERABLE GUIDE CATHETER; VALVE REPAIR Back to Search Results
Catalog Number SGC0301
Device Problem Adverse Event Without Identified Device or Use Problem (2993)
Patient Problem Thrombosis (2100)
Event Date 09/04/2019
Event Type  Injury  
Manufacturer Narrative
Exemption number (b)(4).The customer reported the device was discarded.Investigation is not yet complete.A follow-up will be submitted with all relevant information.
 
Event Description
This is filed to report thrombus.It was reported that this was a mitraclip procedure to treat mitral regurgitation (mr) with a grade of 4.The steerable guiding catheter (sgc) was advanced and the first clip was implanted, reducing mr to 2.Thrombus was seen at the superior vena cava.The activated clotting time (act) was 317 seconds.No additional clips were implanted.The patient was put on medication and on (b)(6) 2019 the patient was doing well.It cannot be determined if the sgc caused or contributed to the thrombus.No additional information was provided.
 
Manufacturer Narrative
Exemption number e2019001-permits numbering sequence to begin with 10000, to avoid duplication of report numbers due to process transition.There may be gaps in numbering for reports submitted during the transition period.The device was not returned for evaluation.A review of the lot history record identified no manufacturing nonconformities issued to the reported lot that would have contributed to this event.The reported patient effect of thrombosis as listed in the mitraclip system instructions for use, is a known possible complication associated with mitraclip procedures.A conclusive cause for the reported patient effect, and the relationship to the product, if any, cannot be determined.There is no indication of a product quality issue with respect to manufacture, design or labeling.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
STEERABLE GUIDE CATHETER
Type of Device
VALVE REPAIR
Manufacturer (Section D)
ABBOTT VASCULAR
26531 ynez rd.
temecula CA 92591 4628
MDR Report Key9111740
MDR Text Key162001469
Report Number2024168-2019-12053
Device Sequence Number1
Product Code DRA
UDI-Device Identifier08717648216824
UDI-Public08717648216824
Combination Product (y/n)N
PMA/PMN Number
K161985
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Type of Report Initial,Followup
Report Date 10/29/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Health Professional
Device Expiration Date07/16/2020
Device Catalogue NumberSGC0301
Device Lot Number90716U250
Was Device Available for Evaluation? No
Initial Date Manufacturer Received 09/04/2019
Initial Date FDA Received09/24/2019
Supplement Dates Manufacturer Received10/17/2019
Supplement Dates FDA Received10/29/2019
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
Patient Outcome(s) Required Intervention;
Patient Age82 YR
-
-