• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: DAVOL INC., SUB. C.R. BARD, INC. -1213643 3DMAX LIGHT; SURGICAL MESH

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

DAVOL INC., SUB. C.R. BARD, INC. -1213643 3DMAX LIGHT; SURGICAL MESH Back to Search Results
Catalog Number 0117310
Medical Device Problem Codes Defective Device (2588); Insufficient Device Problem Information (3190); Migration (4003)
Health Effect - Clinical Codes Adhesion(s) (1695); Hernia (2240); Injury (2348)
Date of Event 11/29/2016
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
At this time no conclusions can be made.The patient's attorney alleges surgical intervention; however, no details have been provided.No lot number has been provided; therefore a review of the manufacturing records is not possible.Should additional information be provided a supplemental emdr will be submitted.Not returned.
 
Event or Problem Description
Attorney alleges that the patient underwent surgery for implant of an unspecified bard/davol 3dmax light mesh on (b)(6) 2016.It is alleged that the patient had subsequent surgical intervention due to the hernia mesh device(s).As reported, the patient is making a claim for an adverse patient outcome against the bard/davol 3dmax light mesh.As reported, the attorney alleges patient experienced emotional distress and the device was defective.
 
Event or Problem Description
Attorney alleges that the patient underwent surgery for implant of an unspecified bard/davol 3dmax light mesh on (b)(6) 2016.It is alleged that the patient had subsequent surgical intervention due to the hernia mesh device(s).As reported, the patient is making a claim for an adverse patient outcome against the bard/davol 3dmax light mesh.As reported, the attorney alleges patient experienced emotional distress and the device was defective.Addendum per additional information provided: (b)(6) 2016 - patient was diagnosed with left recurrent direct and indirect inguinal hernia thereby underwent laparoscopic repair with the implant of 3dmax light mesh.Per operative notes, "the hernia sac was seen in the spermatic cord and another hernia sac medial to the inferior epigastric vessels.A pantaloon hernia, direct and an indirect was seen.The indirect sac was mobilized.A 3dmax light mesh was placed." (b)(6) 2016 - patient was diagnosed with left recurrent direct groin hernia thereby underwent open repair with the implant of synthetic mesh.Per operative notes, "a large direct defect was adherent to the cord.The previous laparoscopic mesh had migrated medially allowing for a direct hernia.The direct hernia sac was reduced and the floor of the inguinal canal was repaired.A synthetic mesh was placed and sutured.".
 
Additional Manufacturer Narrative
At this time no conclusions can be made.The patient's attorney alleges surgical intervention; however, no details have been provided.No lot number has been provided; therefore a review of the manufacturing records is not possible.Addendum: this supplemental emdr is submitted to document additional information provided.Root cause is inconclusive; no conclusion can be made as to the extent to which the implant may have caused or contributed to the patient's postoperative course.Note, the manufacturing date (15-jun-2015) is considered to be a best estimate.Updated fields: a2, a4, b3, b4, b5, b7, d4, g1, g3, g6, h2, h4, h6, h10.Note: section a through f - the information provided by bd represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.H3 other text: not returned.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
3DMAX LIGHT
Common Device Name
SURGICAL MESH
Manufacturer (Section D)
DAVOL INC., SUB. C.R. BARD, INC. -1213643
100 crossings blvd.
warwick RI 02886
Manufacturer (Section G)
BARD SHANNON LIMITED 3005636544
san geronimo industrial park
lot #1, road #3, km 79.7
humacao PR 00791
Manufacturer Contact
andrew topoulos
100 crossings blvd.
warwick, RI 02886
4018258495
MDR Report Key9111999
Report Number1213643-2019-09228
Device Sequence Number19729938
Product Code FTL
UDI-Device Identifier00801741031021
UDI-Public(01)00801741031021(17)200628(10)HUZF0649
Combination Product (Y/N)N
Initial Reporter StateNY
Initial Reporter CountryUS
PMA/510(K) Number
K091659
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2016
Reporter Type Manufacturer
Report Source Other,Consumer
Initial Reporter Occupation Non-Healthcare Professional
Type of Report Initial,Followup
Report Date (Section B) 07/04/2025
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date06/28/2020
Device Catalogue Number0117310
Device Lot NumberHUZF0649
Was Device Available for Evaluation? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 09/04/2019
Supplement Date Received by Manufacturer06/13/2025
Initial Report FDA Received Date09/24/2019
Supplement Report FDA Received Date07/04/2025
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured06/15/2015
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient Age48 YR
Patient SexMale
Patient Weight70 KG
-
-