|
The patient was undergoing a thrombectomy procedure using penumbra system 3max reperfusion catheters (3maxcs).During the procedure, the physician successfully advanced a 3maxc through a penumbra system ace 68 reperfusion catheter (ace68) to the target location.When the 3maxc was being retracted from the ace68, the physician experienced resistance.Upon removal of the 3maxc from the tuohy, the 3maxc fractured into two pieces.The physician then used hemostat forceps to remove the remaining piece of 3maxc from the ace68.Therefore, the 3maxc was completely removed from the procedure.The physician continued the procedure using another 3maxc, however, the same issue occurred.This 3maxc was also removed and was no longer used in the procedure.The procedure was completed using a third 3maxc and the same ace68.There was no report of an adverse effect to the patient.
|