• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ALERE SAN DIEGO ALERE HCG CASSETTE (25MIU/ML) 40T; HCG PREGNANCY TEST

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ALERE SAN DIEGO ALERE HCG CASSETTE (25MIU/ML) 40T; HCG PREGNANCY TEST Back to Search Results
Model Number 92217
Medical Device Problem Code False Positive Result (1227)
Health Effect - Clinical Code No Consequences Or Impact To Patient (2199)
Date of Event 07/05/2017
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
At the time of the event, there was no indication of a reportable malfunction based on information provided and assessed by alere (b)(4).Alere (b)(4) updated reporting decisions and conducted a retrospective review of complaints against the updated reporting decisions.This mdr is a retrospective filing that was identified during this retrospective review activity associated with an observation from an fda inspection conducted in january 2019 at alere (b)(4) (reference eir (b)(4)).The awareness date is based on updated reporting decisions and completion of the retrospective review activity.There is no new or increased trend based on this retrospective review activity.Investigation results: the customer's observation was not replicated in-house with retention and return products of the reported lot.Retention and return products were tested with low concentration hcg urine standards and clinically negative urine samples.All devices showed expected negative results at read time.No false positive results were obtained during in-house testing.Manufacturing batch record review did not uncover any abnormalities.Root cause could not be determined based on the information provided.
 
Event or Problem Description
It was reported that a customer received a false positive result when using an alere hcg cassette.The results confirmed via quant (b)(6) 2017 hcg = <2.39 miu/mlusing vitros 3600.The customer stated their facility reports patients as not pregnant if hcg less than 6.15 miu/ml.No other patient information was received.Troubleshooting occurred with a discussion about the issue and potential factors per the pi, storage, sampling, and technique.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ALERE HCG CASSETTE (25MIU/ML) 40T
Common Device Name
HCG PREGNANCY TEST
Manufacturer (Section D)
ALERE SAN DIEGO
9995 summers ridge rd
san diego CA 92121
MDR Report Key9114347
Report Number2027969-2019-00398
Device Sequence Number2307941
Product Code JHI
UDI-Device Identifier10815845020109
UDI-Public(01)10815845020109(17)180831(10)HCG6090073
Combination Product (Y/N)N
Initial Reporter StateWV
Initial Reporter CountryUS
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source health professional,user faci
Initial Reporter Occupation Administrator/Supervisor
Type of Report Initial
Report Date (Section B) 09/24/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date08/31/2018
Device Model Number92217
Device Lot NumberHCG6090073
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer07/26/2017
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 07/19/2019
Initial Report FDA Received Date09/25/2019
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-