It was reported by a lab director that a patient came into the clinic for contraceptives and a false positive result occurred with the hcg serum/urine test.The patient tested positive on hcg test but the serum quant came back at <2miu/ml.The patient was instructed to come back in a few weeks.A second rapid test also came back as a false positive.No patient outcome provided; unknown if patient received contraceptive treatment.Trouble shooting occurred with a review of storage, technique, and handling - no deviations noted.Tss discussed using only clear urine specimens free of particulates, variance of hcg clearance rates in individuals post pregnancy, and possibility of hama interference.
|