The subject device was returned to olympus medical systems corp.(omsc) for evaluation.The tissue pad of the subject device was worn and partially separated.The manufacturing record was reviewed and found no irregularities.Based on the past similar cases, it was known that the tissue pad was damaged since the user continued to activate output without grasping tissue (including after the tissue already cut).The above device handling has warned in the instruction manual as follows: do not activate output in seal & cut mode while the grasping section is closed without contacting tissue or vessel, or ensuring that tissue is transected.Otherwise, a local increase of the temperature due to a friction between the probe tip and the grasping section may result in various forms of damage in the probe tip and/or the tissue pad, such as premature wear, breakage, deformation, and/or falling off inside the body cavity and/or partial separating.
|
|
During an unspecified laparoscopic procedure, the subject device was used.After one activation, the tissue pad of the subject device was damaged and separated.No further information was provided.On september 24, 2019, olympus medical systems corp.(omsc) found that the tissue pad of the subject device was partially separated.This is the report regarding the separation of the tissue pad.
|