• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ETHICON INC. STLESS STEEL SIL 4X18IN 5 S/A CCS; SUTURE, NONABSORBABLE, STEEL

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ETHICON INC. STLESS STEEL SIL 4X18IN 5 S/A CCS; SUTURE, NONABSORBABLE, STEEL Back to Search Results
Catalog Number M653G
Medical Device Problem Code Break (1069)
Health Effect - Clinical Codes Wound Dehiscence (1154); Blood Loss (2597); Not Applicable (3189)
Date of Event 08/19/2019
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
(b)(4).A manufacturing record evaluation was performed for the finished device al7975 number, and no non-conformance's were identified.To date the device has not been returned.If the device or further details are received at a later date a supplemental medwatch will be sent.Attempts are being made to obtain the following information.To date no response has been provided.If further details are received at a later date a supplemental medwatch will be sent.The patient demographic info: age, weight, bmi at the time of index procedure? what was the name of the initial procedure? what was the date of the initial procedure? can you describe how the steel suture was placed (how many times twisted)? what tissue was the steel suture used on (sternum)? what symptoms did patient present with post op? how many post op days did the patient present with symptoms? can you describe the appearance of the steel suture during second procedure? was the steel suture removed? what was done to manage the post op steel breakage? what is physician¿s opinion as to the etiology of or contributing factors to the breakage? is the steel suture that was removed, be returned, return date, tracking information? do you have any samples of lot al7975 available for evaluation? what is the current status of the patient?.
 
Event or Problem Description
It was reported that the patient underwent sternal suture on an unknown date and suture was used.The suture broke post-operatively.The patient underwent removal of suture and revision surgery on (b)(6) 2019.Additional information has been requested.
 
Additional Manufacturer Narrative
(b)(4).Additional information was requested and the following was obtained: the patient demographic info: age, weight, bmi at the time of index procedure 62 years old.What was the name of the initial procedure? thoracic aortic aneurysm - surgical correction.What was the date of the initial procedure? july 27, 2019 can you describe how the steel suture was placed (how many times twisted)? i do not have this information, as the procedure was performed by medical staff at the surgical center.What tissue was the steel suture used on (sternum)? chest wall suture.What symptoms did patient present with post op? patient who underwent cardiac surgery on (b)(6) 2019, and was transferred today to san francisco hospital in americana on (b)(6) 2019.How many post op days did the patient present with symptoms? patient returned to service on (b)(6) 2019 referred to "americana" (city) for presenting sternal bleeding + sternorrhaphy dehiscence.Was the revision date 7/30/2019? on (b)(6) 2019 the patient underwent chest wall resection.Can you describe the appearance of the steel suture during the second procedure? i do not have this information.Was the steel suture removed? yes, redone every procedure again.What was done to manage the post op steel breakage? image registration.What is physician's opinion as to the etiology of or contributing factors to the breakage? the diagnostic hypotheses mentioned in medical evolution are: sternal bleeding + sternorrhaphy dehiscence.Is the steel suture that was removed, be returned, return date, tracking information? all surgical threads have traceability since the entry of the invoice, with release through internal code in surgical kit and possible monitoring through the internal system of the hospital.Do you have any samples of lot al7975 available for evaluation? unfortunately we have no sample, all threads of this lot were used before detecting the fact.What is the current status of the patient? patient was discharged on september 2, 2019, returning to the referral hospital of the agreement (san francisco de americana) for the end of clinical treatment.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
STLESS STEEL SIL 4X18IN 5 S/A CCS
Common Device Name
SUTURE, NONABSORBABLE, STEEL
Manufacturer (Section D)
ETHICON INC.
p.o. box 151, route 22 west
somerville NJ 08876 0151
MDR Report Key9117757
Report Number2210968-2019-88090
Device Sequence Number2347080
Product Code GAQ
UDI-Device Identifier10705031045477
UDI-Public10705031045477
Combination Product (Y/N)N
Initial Reporter CountryBR
PMA/510(K) Number
K946173
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,foreig
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 08/29/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date01/31/2024
Device Catalogue NumberM653G
Device Lot NumberAL7975
Was Device Available for Evaluation? No
Initial Date Received by Manufacturer 08/29/2019
Supplement Date Received by Manufacturer10/07/2019
Initial Report FDA Received Date09/25/2019
Supplement Report FDA Received Date11/01/2019
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient Age62 YR
-
-