| Model Number |
53401 |
| Medical Device Problem Code |
Thermal Decomposition of Device (1071)
|
| Health Effect - Clinical Codes |
No Patient Involvement (2645); No Known Impact Or Consequence To Patient (2692)
|
| Date of Event |
08/24/2019
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Nova biomedical opened an internal investigation to evaluate the returned docking station, sn (b)(4), and meters, sn (b)(4).
|
| |
|
Event or Problem Description
|
|
On (b)(6) 2019 nova was made aware of an issue regarding nova stat strip docking station, sn (b)(4), sparking and smoking.
|
| |
|
Additional Manufacturer Narrative
|
|
Device investigation: the device history records (dhrs) were reviewed by the senior quality control manager, and indicated that the released product met all specifications.The two meters were inspected and performed as inspected, with no failures found.Disassembly of the docking station showed evidence of overheating.There was also evidence of debris and fluid ingress inside the docking station, which was likely caused by fluid spillage.The safety measures built into the docking station design caused the adaptor board connector to melt instead of catching fire.Since this docking station was manufactured, the adaptor board fuse has been moved to the underside of the docking station board to protect it from fluid ingress, moving the point of power interruption upstream.This will prevent the issue from re-occurring under similar circumstances.As a result of the investigation, the complaint has been confirmed.Based on the evidence of the damage, the root cause has been determined to be improper user handling.
|
| |
|