ALLERGAN (COSTA RICA) INSPIRA TEXTURED SILICONE GEL FILLED BREAST IMPLANT; PROSTHESIS, BREAST, NONINFLATABLE, INTERNAL, SILICONE GEL-FILLED
|
Back to Search Results |
|
| Model Number |
TSLP-205 |
| Medical Device Problem Code |
Adverse Event Without Identified Device or Use Problem (2993)
|
| Health Effect - Clinical Code |
Pain (1994)
|
| Date of Event |
07/29/2019
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
A review of the device history record has been initiated.If any new, changed or corrected information is noted, a supplemental medwatch will be submitted.The event of pain is a physiological complication and analysis of the device generally does not assist allergan in determining a probable cause for this event.Further information from the reporter regarding event, product, or patient details has been requested.No additional information is available at this time.Reason for reoperation: pain.This is a known potential adverse event addressed in the product labeling.
|
| |
|
Event or Problem Description
|
|
Healthcare professional reported left side "tightening, pain, and lumps." device has been explanted.
|
| |
|
Additional Manufacturer Narrative
|
|
Device evaluation: visual analysis of the returned device identified: deformation, fold creases, air voids between shell and gel.And was observed after autoclave cycle: cloudy gel.A weight test of the explanted material was verified and it was within specification.Based on the device analysis the final assessment is: no observations found related with the complaint reported by the physician.Additional, changed, and/or corrected data: d.10, g.1., h.3., h.6.
|
| |
|
Event or Problem Description
|
|
Healthcare professional reported left side "tightening, pain, and lumps." device has been explanted.
|
| |
|