|
Reportable based on device analysis completed on (b)(6) 2019.It was reported that the main coil was stretched.The target area was located in a moderately tortuous internal iliac artery.During procedure, after an imager bern catheter was inserted via cross over approach, a 6mm x 20cm and a 4mm x 6cm interlock coil were deployed.A 3mm x 4cm.035 interlock coil was used for embolization.Subsequently, the 3mm x 4cm.035 interlock was inserted; however, the coil became unraveled/stretched.The device was removed from the patient's body and imaging was performed.The blood flow was blocked, thus, the procedure was ended.There were no patient complications reported.However, device analysis revealed a detached interlocking arm.
|