• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BD MEDICAL - DIABETES CARE BD SYRINGE 0.5ML 29G 12.7MM; PISTON SYRINGE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BD MEDICAL - DIABETES CARE BD SYRINGE 0.5ML 29G 12.7MM; PISTON SYRINGE Back to Search Results
Catalog Number 323003
Medical Device Problem Code Device Contamination with Chemical or Other Material (2944)
Health Effect - Clinical Code No Known Impact Or Consequence To Patient (2692)
Date of Event 09/06/2019
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
A device evaluation is anticipated, but has not yet begun.Upon completion of the investigation, a supplemental report will be filed.
 
Event or Problem Description
It was reported that foreign matter was discovered with a bd syringe 0.5ml 29g 12.7mm.This occurred on 2 separate occasions during use, however, the date/time information is unknown.The following information was provided by the initial reporter: it was reported that black foreign matter was found within the syringe.
 
Additional Manufacturer Narrative
H.6.Investigation summary: no samples (including photos) were returned therefore the complaint could not be confirmed and the root cause is undetermined.Complaints received for this device and reported condition will continue to be tracked and trended.If samples are received in the future the complaint will be reopened for further investigation.A review of the device history record was completed for batch#: 6294537.All inspections and challenges were performed per the applicable operations qc specifications.There were zero (0) notifications noted that pertained to the complaint.As no samples and/or photo(s) were received the investigation concluded: unconfirmed: bd was not able to duplicate or confirm the customer¿s indicated failure as no samples or photos were returned.Based on the above, no additional investigation and no capa is required at this time.H3 other text : see section h.10.
 
Event or Problem Description
It was reported that foreign matter was discovered with a bd syringe 0.5ml 29g 12.7mm.This occurred on 2 separate occasions during use, however, the date/time information is unknown.The following information was provided by the initial reporter: it was reported that black foreign matter was found within the syringe.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
BD SYRINGE 0.5ML 29G 12.7MM
Common Device Name
PISTON SYRINGE
Manufacturer (Section D)
BD MEDICAL - DIABETES CARE
1329 west highway 6
holdrege NE 68949
MDR Report Key9119313
Report Number1920898-2019-01036
Device Sequence Number343827
Product Code FMF
UDI-Device Identifier00382903230020
UDI-Public00382903230020
Combination Product (Y/N)N
Initial Reporter CountryCA
PMA/510(K) Number
EXEMPT
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source consumer,foreign,other
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 10/14/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Lay User/Patient
Device Expiration Date10/31/2021
Device Catalogue Number323003
Device Lot Number6294537
Initial Date Received by Manufacturer 09/10/2019
Supplement Date Received by Manufacturer09/10/2019
Initial Report FDA Received Date09/25/2019
Supplement Report FDA Received Date10/14/2019
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Other;
-
-