Model Number 42856
Medical Device Problem Code Appropriate Device Problem Term/Code Not Available (3191)
Health Effect - Clinical Code No Known Impact Or Consequence To Patient (2692)
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
If information is provided in the future, a supplemental report will be issued.
 
Event or Problem Description
It was reported that the valve had poor drainage.The valve was replaced, and the new valve was working normally.
 
Additional Manufacturer Narrative
The pl setting on the returned valve was observed to be between 2.5 and 0.5.The laboratory personnel could not adjust the pl setting on the valve prior to or after flushing.Therefore, the valve could not be tested for valve flux, patency, leakage, siphon, reflux, pressure/flow and pre-implantation tests.It was observed that water was able to flow through the valve during flushing.Biological debris were observed on the interior of the valve.The instructions for use cautions, ¿shunt obstruction may occur in any of the components of the shunt system.The system may become occluded internally due to tissue fragments, blood clots, tumor cell aggregates, bacterial colonization or other debris.¿ all valves are 100% tested at the time of manufacture.If information is provided in the future, a supplemental report will be issued.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
STRATA II VALVE, SMALL
Common Device Name
SHUNT, CENTRAL NERVOUS SYSTEM AND COMPONENTS
Manufacturer (Section D)
MEDTRONIC NEUROSURGERY
125 cremona drive
goleta CA 93117
MDR Report Key9120284
Report Number2021898-2019-00346
Device Sequence Number1924452
Product Code JXG
UDI-Device Identifier00643169975828
UDI-Public00643169975828
Combination Product (Y/N)N
Initial Reporter CountryCH
PMA/510(K) Number
K042465
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2019
Reporter Type Manufacturer
Report Source foreign,health professional
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 12/26/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date11/30/2020
Device Model Number42856
Device Catalogue Number42856
Device Lot NumberE41719
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer10/25/2019
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 09/24/2019
Supplement Date Received by Manufacturer12/19/2019
Initial Report FDA Received Date09/25/2019
Supplement Report FDA Received Date12/26/2019
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Hospitalization; Required Intervention;
Patient Weight51
-
-