| Catalog Number |
523393 |
| Medical Device Problem Code |
Adverse Event Without Identified Device or Use Problem (2993)
|
| Health Effect - Clinical Codes |
Foreign Body Reaction (1868); Discomfort (2330); Joint Disorder (2373); No Code Available (3191)
|
| Date of Event |
01/12/2015
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
(b)(4).Initial reporter occupation: lawyer.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
|
| |
|
Event or Problem Description
|
|
Ppf and medical records received.There were no new allegation reported from ppf.After review of medical records, patient was revised to address chronic discomfort and bursitis versus abductor tendon tears but operative notes reported that abductor tendon appear to be intact.Revision note reported metallosis and the posterior aspect of the femoral neck was impinging on the retaining ring on the acetabular component.Doi: (b)(6) 2013; dor: (b)(6) 2015; (left hip) third revision.Please see (b)(4) for the first revision.Please see (b)(4) for the second revision.
|
| |
|
Additional Manufacturer Narrative
|
|
Depuy synthes is submitting this report pursuant to the provision depuy synthesof 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.Product complaint # = > (b)(4).Investigation summary = > no device associated with this report was received for examination.A worldwide complaint database search found no other related reported incidents against the provided product code/lot number combination since release for distribution.The information received will be retained for potential series investigations if triggered by trend analysis, post market surveillance, or other events within the quality system.Device history lot = > null.Device history batch = > null.Device history review = > null.If information is obtained that was not available for the initial medwatch, a follow-up medwatch, a follow-up medwatch will be filed as appropriate.
|
| |
|
Search Alerts/Recalls
|