K2M, INC. SET SCREW; ATR; THORACOLUMBOSACRAL PEDICLE SCREW SYSTEM Back to Search Results
Model Number 101-10001
Medical Device Problem Code Adverse Event Without Identified Device or Use Problem (2993)
Health Effect - Clinical Codes Injury (2348); No Code Available (3191)
Date of Event 08/08/2019
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Return status of the device/devices is currently unknown.
 
Event or Problem Description
A physician reported four denali set screws backed-out post-operatively leading to revision surgery.This record captures the third of the four set screws.
 
Additional Manufacturer Narrative
Correction to d.3.And g.1.: updated from 'stryker spine- leesburg' to 'k2m, inc.' upon review of provided x-rays, it was observed that bilateral screws at the caudal end of the construct disengaged from their associated rods.There was no issue found with the third denali set screw.Therefore, this device is concomitant.
 
Event or Problem Description
A physician reported four denali set screws backed-out post-operatively leading to revision surgery.This record captures the third of the four set screws.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SET SCREW; ATR
Common Device Name
THORACOLUMBOSACRAL PEDICLE SCREW SYSTEM
Manufacturer (Section D)
K2M, INC.
600 hope parkway se
leesburg VA 20175
MDR Report Key9120803
Report Number3004774118-2019-00100
Device Sequence Number19526895
Product Code NKB
UDI-Device Identifier10888857003842
UDI-Public10888857003842
Combination Product (Y/N)N
Initial Reporter CountryCO
PMA/510(K) Number
K141873
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2019
Device Explanted Year2019
Reporter Type Manufacturer
Report Source health professional
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 03/17/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number101-10001
Device Catalogue Number101-10001
Device Lot NumberJEWU
Was Device Available for Evaluation? No
Initial Date Received by Manufacturer 08/27/2019
Supplement Date Received by Manufacturer02/28/2020
Initial Report FDA Received Date09/25/2019
Supplement Report FDA Received Date03/17/2020
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
-
-