• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: LIVANOVA DEUTSCHLAND HEATER-COOLER SYSTEM 3T; CONTROLLER, TEMPERATURE, CARDIOPULMONARY BYPASS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

LIVANOVA DEUTSCHLAND HEATER-COOLER SYSTEM 3T; CONTROLLER, TEMPERATURE, CARDIOPULMONARY BYPASS Back to Search Results
Model Number 16-02-80
Device Problem Melted (1385)
Patient Problem No Patient Involvement (2645)
Event Date 08/27/2019
Event Type  malfunction  
Manufacturer Narrative
There was no patient involvement.The heater-cooler 16-02-80 is not distributed in the usa, but it is similar to heater-cooler 16-02-85, which is distributed in the usa (510(k) number: k052601).Livanova (b)(4) manufactures the heater-cooler system 3t.The incident occurred in (b)(6).A review of the dhr did not identify any deviations or non-conformities relevant to the reported issue.If any additional information pertinent to the reported event is received, it will be provided in a supplemental report.
 
Event Description
Livanova (b)(4) received a report that the main plug of a heater-cooler system 3t could not be removed from the socket because the plug plastic had melted and glued to the wall socket.An electrician could remove the plug by force eventually.There was no patient involvement.
 
Manufacturer Narrative
H.10: since the plug had been cut-off on site, it was not made available for further investigation.The power cable will be replaced in order to solve the issue.
 
Event Description
See initial report.
 
Event Description
See initial report.
 
Manufacturer Narrative
H.10: as stated in the previous report, the part was not made available for further investigation.However, the investigation carried out on similar cases from the same hospital is most likely applicable also to this specific case.During the investigation performed by the manufacturer it was confirmed that one pin of the mains connector was exposed to large heat.The heat effect took place with high probability from the outside over the pin to the inside.From an electrical point of view, the plug and the cable were in good conditions, thus, no damage except for the fused point on the plug.The cause of the damage was very likely a faulty hospital wall outlet.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
HEATER-COOLER SYSTEM 3T
Type of Device
CONTROLLER, TEMPERATURE, CARDIOPULMONARY BYPASS
Manufacturer (Section D)
LIVANOVA DEUTSCHLAND
lindberghstr. 25
munich
MDR Report Key9122029
MDR Text Key195714581
Report Number9611109-2019-00749
Device Sequence Number1
Product Code DWC
Combination Product (y/n)N
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type health professional
Remedial Action Other
Type of Report Initial,Followup,Followup
Report Date 01/08/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received09/26/2019
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number16-02-80
Was the Report Sent to FDA? No
Date Manufacturer Received12/09/2019
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-