• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MEDTRONIC EATONTOWN GRAFTON DBM; BONE GRAFTING MATERIAL, HUMAN SOURCE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MEDTRONIC EATONTOWN GRAFTON DBM; BONE GRAFTING MATERIAL, HUMAN SOURCE Back to Search Results
Catalog Number T43110INT
Medical Device Problem Code Device Appears to Trigger Rejection (1524)
Health Effect - Clinical Code Discomfort (2330)
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Product was not returned to the manufacturer for evaluation.Therefore, we are unable to determine the definitive cause of the reported event.Hcps review based on donor records: after reviewing the donor records, and the complaint of the surgeon, the hcps see no objective evidence that the graft was the cause of the patient¿s ¿relapse¿.The fact that there was ¿no post-op infection, or immunological or inflammatory response¿ reinforces their conclusion.Manufacturing assessment: no deviations from established procedures were noted that may have increased the risk of contamination/cross contamination.All supplies and reagents utilized were inspected upon receipt and prior to recovery.The supplies and reagents were stored as per packaging or manufacturer recommendation in a temperature controlled environment.Products from multiple manufacturers were implanted during the procedure.Although it is unknown if any of the devices contributed to the reported event, we are filing this mdr for notification purposes.If information is provided in the future, a supplemental report will be issued.
 
Event or Problem Description
It was reported that the patient underwent laminectomy and posterior arthrodesis at l4-l5.During the surgery, the surgeon placed 10 cc putty into the patient.No cage was used; the surgeon only put screws and rod manufactured by medi implants.No intra-operative complications were reported.Approximately a week after the surgery, reportedly, the patient had a relapse and had low levels of saturation among other symptoms.The patient reported experiencing a rare feeling.The surgeon planned a revision surgery due to this event.On (b)(6) 2019, the revision surgery was performed.When the wound was opened, it was observed that the graft had a black-gray color and had an abnormal appearance.The surgeon then commented that the graft was bad and that the patient's body could have rejected the graft.Hence, everything implanted during the initial surgery was completely explanted.The items that were explanted included the reported graft, the screws and the rod.The patient had to be hospitalized due to this event.There was no post-op infection, or immunological or inflammatory response reported.Patient issue has been reported to be resolved.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
GRAFTON DBM
Common Device Name
BONE GRAFTING MATERIAL, HUMAN SOURCE
Manufacturer (Section D)
MEDTRONIC EATONTOWN
201 industrial way west
eatontown NJ 07724
MDR Report Key9122085
Report Number2246640-2019-00008
Device Sequence Number2597115
Product Code NUN
UDI-Device Identifier00643169965003
UDI-Public00643169965003
Combination Product (Y/N)N
Initial Reporter CountryCO
PMA/510(K) Number
K051188
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2019
Device Explanted Year2019
Reporter Type Manufacturer
Report Source company representative,foreig
Initial Reporter Occupation Physician
Type of Report Initial
Report Date (Section B) 09/26/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date01/16/2022
Device Catalogue NumberT43110INT
Was Device Available for Evaluation? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 08/27/2019
Initial Report FDA Received Date09/26/2019
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Hospitalization; Required Intervention;
-
-