| Catalog Number |
0375941012 |
| Medical Device Problem Code |
Excessive Heating (4030)
|
| Health Effect - Clinical Code |
Burn(s) (1757)
|
| Date of Event |
08/26/2019
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
Additional information will be provided once the investigation has been completed.
|
| |
|
Event or Problem Description
|
|
It was reported that the device burned the patient.
|
| |
|
Additional Manufacturer Narrative
|
|
This complaint investigation was closed based on the device not received, therefore the reported failure mode was not confirmed.In the event that the device is received, the complaint will be reopened and the investigation will be updated with new results.The reported failure mode will be monitored for future reoccurrence alleged failure: patient burn.Probable root cause: - manufacturing defect (grease application) - high friction due to components interaction - surface not smooth - poor/lack of suction - clogged shaver blade preventing fluid suction and cooling.
|
| |
|
Event or Problem Description
|
|
It was reported that the device burned the patient.
|
| |
|