|
The referenced device will not be returned to olympus, therefore, the exact cause of the reported event cannot be confirmed.The oem conducted a dhr review; manufacturing and quality control review was performed for the affected lot or serial number without showing any non-conformities or deviations regarding the described issue.Based on similar reported complaints, the reported event can potentially occur if the electrode comes into unintended contact with other metal parts, e.G., surgical instruments while the high-frequency output was activated.As a preventive measure, the instructions for use contains several warning and caution statements to prevent damage to the electrode: "visually inspect the instrument prior to use.Warped electrodes, defective insulation and broken, cracked, or irregular cutting loops represent a danger for both the patient and the surgeon and must not be used.If the instrument is damaged or does not function properly, replace it." if additional information is received it will be reviewed and a supplemental report will be filed.
|