• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: DAVIS & GECK CARIBE LTD BIOSYN; SUTURE, ABSORBABLE, SYNTHETIC, POLYGLYCOLIC ACID

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

DAVIS & GECK CARIBE LTD BIOSYN; SUTURE, ABSORBABLE, SYNTHETIC, POLYGLYCOLIC ACID Back to Search Results
Model Number SM-632
Medical Device Problem Code Detachment of Device or Device Component (2907)
Health Effect - Clinical Code No Consequences Or Impact To Patient (2199)
Date of Event 08/30/2019
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
If information is provided in the future, a supplemental report will be issued.
 
Event or Problem Description
According to the reporter, during a procedure, when the first packet was opened, the needle and the thread were found to be disengaged, so the product could not be used.Another product was used to complete the case.There was no patient injury.
 
Additional Manufacturer Narrative
Evaluation summary: post market vigilance (pmv) led an evaluation of one device the visual inspection of the sample noted one suture and one needle.The needle was returned disengaged from the suture.Upon microscopic inspection of the needle, normal crimp and swage marks were noted.As no sealed samples were returned, a needle attachment test could not be performed.Records from each manufacturing lot are thoroughly reviewed to ensure that products are released meeting all quality release specifications at the time of manufacture.A definitive root cause could not be determined with regard to the reported condit medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
BIOSYN
Common Device Name
SUTURE, ABSORBABLE, SYNTHETIC, POLYGLYCOLIC ACID
Manufacturer (Section D)
DAVIS & GECK CARIBE LTD
zona franca de san isidro
santo domingo 18750
DO  18750
MDR Report Key9126717
Report Number9612501-2019-01825
Device Sequence Number2614561
Product Code GAM
UDI-Device Identifier20884521033167
UDI-Public20884521033167
Combination Product (Y/N)N
Initial Reporter CountryJA
PMA/510(K) Number
K000037
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source foreign,health professional,u
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 12/23/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date09/30/2023
Device Model NumberSM-632
Device Catalogue NumberSM-632
Device Lot NumberD8K2526FY
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer10/22/2019
Initial Date Received by Manufacturer 09/03/2019
Supplement Date Received by Manufacturer12/19/2019
Initial Report FDA Received Date09/26/2019
Supplement Report FDA Received Date12/23/2019
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-