MAUDE Adverse Event Report: REFURB, SITE-RITE 8; SYSTEM, IMAGING, PULSED ECHO, ULTRASONIC

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

REFURB, SITE-RITE 8; SYSTEM, IMAGING, PULSED ECHO, ULTRASONIC Back to Search Results
Catalog Number 9770501R
Medical Device Problem Codes Loss of Power (1475); Battery Problem (2885)
Health Effect - Clinical Code No Consequences Or Impact To Patient (2199)
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
The device has not been received by the manufacturer, at this time, for evaluation.A history review of serial number (b)(4) showed no other similar product complaint(s) from this serial number.
 
Event or Problem Description
Per biomed: the battery will not hold a charge.Shuts down almost as soon as it's unplugged.
 
Additional Manufacturer Narrative
H11:section a through f - the information provided by bd represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.The following were reviewed as part of this investigation: patient severity, applicable previous investigation(s), sample(if available), applicable fmea documents, labeling and applicable manufacture records.Based on a review of this information, the following was concluded: the device was returned to the service facility for evaluation.During evaluation, the reported issue of battery will not hold a charge.Shuts down almost as soon as it's unplugged was confirmed.The scanner was inspected and it would not charge pass 0% and when unplugged from the ac adapter it would shut down.The root cause of the failure was identified as the internal battery failed.The device was serviced, tested and returned to the customer.A history review of serial number (b)(6) showed no other similar product complaint(s) from this serial number.
 
Event or Problem Description
Per biomed: the battery will not hold a charge.Shuts down almost as soon as it's unplugged.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
REFURB, SITE-RITE 8
Common Device Name
SYSTEM, IMAGING, PULSED ECHO, ULTRASONIC
MDR Report Key9126943
Report Number3006260740-2019-02850
Device Sequence Number14036716
Product Code IYO
UDI-Device Identifier10801741104135
UDI-Public(01)10801741104135
Combination Product (Y/N)N
Initial Reporter StateTX
Initial Reporter CountryUS
PMA/510(K) Number
K152554
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source user facility
Initial Reporter Occupation Biomedical Engineer
Remedial Action Notification
Type of Report Initial,Followup
Report Date (Section B) 04/07/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Catalogue Number9770501R
Device Lot NumberN/A
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer09/27/2019
Event Location Hospital
Initial Date Received by Manufacturer 09/06/2019
Supplement Date Received by Manufacturer03/30/2020
Initial Report FDA Received Date09/26/2019
Supplement Report FDA Received Date04/08/2020
Was Device Evaluated by Manufacturer? (Y/N) Yes
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Reuse
If action reported to FDA under 21 USC 360i(g), list
FDA-assigned Recall Number or include a statement
2243072-10/11/2019-013-C
Patient Sequence Number1
-
-