Catalog Number 284004
Device Problem Suction Failure (4039)
Patient Problem Not Applicable (3189)
Event Date 05/15/2019
Event Type  malfunction  
Manufacturer Narrative
Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.Udi: (b)(4) as of this date, the device has not been returned for evaluation; therefore, the reported condition cannot be confirmed and/or duplicated.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.This report is being filed as required under mitek's corrective and preventative actions (capa) to file usa fda mdr missed malfunctions.
 
Event Description
It was reported by the affiliate in france that during an unspecified surgical procedure, it was observed that the fms vue ii pump short circuited with the fms vue connect interface cable.This resulted in a defective of the interface cable which was beeping continuously.It was also reported that the interface cable could not detect the shaver to generate the suction by the pump.There was no patient impact but there was a two minute delay in surgery.There was a delay in the surgical procedure.It was not reported if there was a spare device available for use to complete the surgery.There was patient involvement.There were no reports of injuries, medical intervention or prolonged hospitalization.All available information has been disclosed.If additional information should become available, a supplemental medwatch will be submitted accordingly.
 
Manufacturer Narrative
Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.If the information is unknown, not available or does not apply, the section/field of the form is left blank.Investigation summary: the device was received and evaluated at the service center.Neither the fault reported by the customer could be verified nor another fault was found with the device upon evaluation.The device was cleaned, tested and found to be fully functional.Since the reported condition is not confirmed,the root cause for the reported failure cannot be determined.A manufacturing record evaluation was performed for the finished device (serial number : (b)(4)), and no non-conformances were identified.At this point in time, no corrective action is required, and no further action is warranted.Depuy mitek will continue to track any related complaints within this device family as a means of monitoring the extent with which this complaint is observed in the field.Device history: a manufacturing record evaluation was performed for the finished device (serial number : (b)(4)), and no non-conformances were identified.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Manufacturer Narrative
Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.If the information is unknown, not available or does not apply, the section/field of the form is left blank.The device was received, the investigation is in progress, no conclusion could be drawn at the time of filing this report.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
FMS VUE PUMP-SHAVER BOX
Type of Device
DISTENSION UNIT, FLUID, ARTHROSCOPIC
Manufacturer (Section D)
MEDOS INTERNATIONAL SàRL
chemin-blanc 38
le locle 02400
SZ  02400
MDR Report Key9129013
MDR Text Key206779644
Report Number1221934-2019-58642
Device Sequence Number1
Product Code HRX
UDI-Device Identifier10886705027934
UDI-Public10886705027934
Combination Product (y/n)N
PMA/PMN Number
K171237
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Type of Report Initial,Followup,Followup
Report Date 05/15/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue Number284004
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer06/05/2019
Initial Date Manufacturer Received 05/15/2019
Initial Date FDA Received09/27/2019
Supplement Dates Manufacturer Received12/03/2019
12/05/2019
Supplement Dates FDA Received12/04/2019
12/06/2019
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-