| Medical Device Problem Codes |
Electrical /Electronic Property Problem (1198); Moisture Damage (1405)
|
| Health Effect - Clinical Code |
No Consequences Or Impact To Patient (2199)
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Device evaluation: the device has been returned and evaluated by product analysis on 05-sep-2019 with the following findings: a visual inspection of the pump revealed the battery compartment was cracked.Moisture damage was observed in the battery compartment.Leak testing revealed a battery compartment leak.The returned battery cap was able to fit securely, however the pump failed to power on.Further testing was unable to be performed due to no power.The pump was opened; moisture damage was found on the printed circuit board.Animas has conducted a review of the device history record for this pump and confirmed that it was operating within required specifications at the time of release.Report late due to transition from vmsr program.
|
| |
|
Event or Problem Description
|
|
On 13-jul-2019, the reporter contacted animas, alleging a power (moisture ingress) issue.There was no indication that the product caused or contributed to an adverse event.This complaint is being reported because the user may be unaware that the pump has lost power, leading to under delivery.
|
| |
|