|
No damages or irregularities were found with the introducer sheath.The concerto pushwire was found broken at the break indicator with the release wire also broken.The proximal segment (actuator interface, positive load indicator, and part of the coupler tube) was not returned for analysis.This is indicative that a manual detachment was attempted at this location.The concerto pushwire was found to be bent at ~6.0cm, ~39.1cm, ~101.6cm and ~171.9cm from the proximal end of the pushwire.The coin was found retracted from the lumen stop.The shield coil was present and slightly stretched with the implant coil already detached and not returned for analysis.The concerto implant coil tip od (outer diameter) could not be measured as it was not returned.The introducer sheath id (inner diameter) was measured and found to be 0.0190¿ (0.4826mm) which is within specification (specification: 0.47mm +0.05/-0.00).No other anomalies were observed.Based on the device analysis and reported information, the customer¿s report of ¿coil resistance/stuck in sheath¿ could not be confirmed as the device was returned with the implant coil already detached.There is evidence that a manual detachment was attempted at the break indicator and the coin was retracted from the lumen stop, detaching the implant coil.The pushwire was found bent throughout the length of the pushwire, indicative of high patient tortuosity.It is also possible that damage to the pushwire occurred during return shipment to medtronic for analysis as the device came back without its protective dispenser coil.The shield coil was found stretched, however the cause could not be determined.It is possible the shield coil became damaged due to attempts to advance or retract the pushwire against the reported resistance or during return shipping to medtronic for analysis.There is no malfunction of the pushwire or non-conformance to specifications that would contribute towards ¿coil resistance/stuck in sheath¿ as the implant coil was not returned for analysis, any contribution of the implant coil towards the failure could not be determined.There is no indication that the event is related to a manufacturing issue.If information is provided in the future, a supplemental report will be issued.
|