As received, the new set was removed from the package.The set was hydrostatic pressure leak tested, with the clave in the activated position, according to product specifications, and no leaks were found anywhere along the fluid path.The original set which generated the complaint was not returned, nor was the mating device.The reported complaint could not be confirmed.A device history (dhr) of the reported lot and relevant commodities were reviewed, and no non-conformances were found that would have contributed to the reported complaint.
|