| Model Number |
FHC-102 |
| Medical Device Problem Code |
False Positive Result (1227)
|
| Health Effect - Clinical Code |
No Information (3190)
|
| Date of Event |
01/30/2019
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
At the time of the event, there was no indication of a reportable malfunction based on information provided and assessed by alere (b)(4).Alere (b)(4) updated reporting decisions and conducted a retrospective review of complaints against the updated reporting decisions.This mdr is a retrospective filing that was identified during this retrospective review activity associated with an observation from an fda inspection conducted in january 2019 at alere (b)(4) (reference eir (b)(4)).The awareness date is based on updated reporting decisions and completion of the retrospective review activity.There is no new or increased trend based on this retrospective review activity.Investigation results: the customer's observation was not replicated in-house with retention and return products.Retention and return devices were tested with in-house clinical hcg negative urine samples.All hcg results were negative at read time and met qc specifications.No false positive results were obtained during in-house testing.Manufacturing batch record review did not uncover any abnormalities.The customer did not provide patient information or information regarding the testing technique.Because return specimen was not available, further investigation was not pursued.A root cause could not be determined based on the information provided.
|
| |
|
Event or Problem Description
|
|
It was reported by the distributor that false positives are occurring with the hcg urine cassette tests at the obgyn facility.Although requested, no further information has been received and no adverse event has been reported.
|
| |
|
Search Alerts/Recalls
|