Model Number MT20649-2
Device Problem Unexpected Shutdown (4019)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 08/25/2019
Event Type  malfunction  
Manufacturer Narrative
(b)(4).
 
Event Description
It was reported that a unexpected receiver shut-down occurred.No product or data was provided for evaluation.Confirmation of the allegation and a root cause could not be determined.No injury or medical intervention was reported.
 
Manufacturer Narrative
(b)(4).
 
Event Description
It was reported that unexpected receiver shutdown occurred.The product was evaluated.The device was visually inspected and it passed.The receiver failed to charge and reboot due to no power.The receiver downloaded by using a known good battery.The log was downloaded and reviewed finding firmware errors.The receiver case was opened and the internal inspection detected battery damage.The allegation was confirmed.The root cause was receiver damaged battery.No injury or medical intervention was reported.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
DEXCOM G4 PLATINUM CONTINUOUS GLUCOSE MONITORING SYSTEM
Type of Device
CONTINUOUS GLUCOSE MONITOR
Manufacturer (Section D)
DEXCOM, INC.
6340 sequence drive
san diego CA 92121
Manufacturer (Section G)
DEXCOM, INC.
6340 sequence drive
san diego CA 92121
Manufacturer Contact
ashley spoto
6340 sequence drive
san diego, CA 92121
8582000200
MDR Report Key9131370
MDR Text Key163768598
Report Number3004753838-2019-69222
Device Sequence Number1
Product Code MDS
Combination Product (y/n)N
Reporter Country CodeSI
PMA/PMN Number
P120005
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type distributor,foreign
Reporter Occupation Other
Type of Report Initial,Followup
Report Date 09/27/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Lay User/Patient
Device Model NumberMT20649-2
Device Catalogue NumberSTK-GL-112
Device Lot Number5223605
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer11/01/2019
Is the Reporter a Health Professional? No
Initial Date Manufacturer Received 08/28/2019
Initial Date FDA Received09/27/2019
Supplement Dates Manufacturer Received11/09/2019
Supplement Dates FDA Received12/04/2019
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured12/15/2016
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-