According to the reporter, the unit had low readings that did not correlate to patient's blood gasses.It was stated that patient was getting normal sat readings and abruptly dropped to 60s and 70s which led to over treating patient.They swapped the sensors, cables and multi measurement servers with no change.The spo2 was low compared to abg's using other device.There was no patient harm.
|
|
Additional information: d10, g4, h3, h6 h3.Evaluation summary: one device was received for evaluation.The reported event was not confirmed.The device met specifications as it was made available for evaluation.It was noted that there was no failure or defect of the product.There was no cause of failure since the investigation found the product to function normally.No new formal investigation is required since there was no fault found with the product.Information has been added to the database and trends will continue to be monitored.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
|