• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: GE MEDICAL SYSTEMS (CHINA) CO., LTD (WUXI) CARESTATION 620 A1; ANESTHESIA GAS MACHINE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

GE MEDICAL SYSTEMS (CHINA) CO., LTD (WUXI) CARESTATION 620 A1; ANESTHESIA GAS MACHINE Back to Search Results
Model Number 1012-9620-000
Device Problem Incorrect, Inadequate or Imprecise Result or Readings (1535)
Patient Problem No Patient Involvement (2645)
Event Date 07/03/2019
Event Type  malfunction  
Manufacturer Narrative
A ge healthcare service representative performed a checkout of the system and confirmed the reported issue.The sensor interface board was replaced to resolve the reported issue.No report of patient involvement.The initial reporter is located outside the u.S., and therefore this information is not provided due to country privacy laws.Date received by manufacturer: this mdr malfunction event was previously eligible for the voluntary malfunction summary reporting (vmsr) program.As of 9/16/2019 notification from fda, this product code is no longer eligible for vmsr.According to the notification, ge healthcare must submit individual reports within 30 calendar days of receiving the notification.Therefore, this event is being reported as an individual mdr report due 10/16/2019.
 
Event Description
The hospital reported a flow sensor error preventing mechanical ventilation.There was no report of patient involvement.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CARESTATION 620 A1
Type of Device
ANESTHESIA GAS MACHINE
Manufacturer (Section D)
GE MEDICAL SYSTEMS (CHINA) CO., LTD (WUXI)
no. 19 changjiang road
national hi-tech dev. zone
wuxi 21402 8
CH  214028
Manufacturer Contact
john szalinski
MDR Report Key9132333
MDR Text Key188266630
Report Number9710602-2019-00085
Device Sequence Number1
Product Code BSZ
UDI-Device Identifier00840682103985
UDI-Public010084068210398521
Combination Product (y/n)N
PMA/PMN Number
K151570
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type foreign,health professional,u
Type of Report Initial
Report Date 09/28/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received09/28/2019
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number1012-9620-000
Was Device Available for Evaluation? Yes
Was the Report Sent to FDA? No
Date Manufacturer Received07/03/2019
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured05/07/2018
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage N
Patient Sequence Number1
-
-