| Catalog Number |
155401120 |
| Medical Device Problem Code |
Adverse Event Without Identified Device or Use Problem (2993)
|
| Health Effect - Clinical Codes |
Foreign Body Reaction (1868); Inflammation (1932); Pain (1994); Discomfort (2330); Test Result (2695); No Code Available (3191)
|
| Date of Event |
08/21/2014
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
Product complaint # (b)(4).Initial reporter occupation: lawyer.(b)(4).If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
|
| |
|
Event or Problem Description
|
|
Litigation alleges that patient experienced pain, discomfort, and inflammation in thigh and groin.Patient also experienced a popping and snapping sensation in hip joint when walking or moving to and from a sitting position.In addition to what were previously alleged, ppf alleges elevated metal ions, metal wear, and metallosis.Added stem due to alleged elevated metal ions.Doi: (b)(6) 2007.Dor: (b)(6) 2014, (right hip).
|
| |
|
Additional Manufacturer Narrative
|
|
Product complaint # (b)(4).Investigation summary = > no device associated with this report was received for examination.The information received will be retained for potential series investigations if triggered by trend analysis, post market surveillance, or other events within the quality system.Depuy considers the investigation closed.Should additional information be received, the information will be reviewed and the investigation will be re-opened as necessary.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
|
| |
|
Search Alerts/Recalls
|