EXACTECH, INC. EQUINOXE; 36MM GLEN TRL Back to Search Results
Model Number 320-01-36
Health Effect - Clinical Code No Known Impact Or Consequence To Patient (2692)
Date of Event 09/11/2019
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Pending evaluation.Concomitant medical device(s): reverse tray (cat# 320-10-00 / sn# (b)(4)).Humeral liner (cat# 320-36-13 / sn# (b)(4)).Torque defining screw kit (cat# 320-20-00 / sn# (b)(4)).Reverse tray adapter plate (cat# 320-10-10 / sn# (b)(4)).Torgue defining screw kit (cat# 320-20-00 / sn# (b)(4)).Rev locking screw (cat# 320-15-05 / sn# (b)(4)).Compress lck cap (cat# 320-20-18 / sn# (b)(4)).Compress lck cap (cat# 320-20-26 / sn# (b)(4)).Compress lck cap (cat# 320-20-22 / sn# (b)(4)).Compress lck cap (cat# 320-20-18 / sn# (b)(4)).Compress lck cap (cat# 320-20-30 / sn# (b)(4)).
 
Event or Problem Description
It was reported that the patient had gone through a shoulder reverse on the (b)(6) of (b)(6) 2019, due to baseplate loosening and instability on the glenoid the shoulder was revised.
 
Additional Manufacturer Narrative
Section h10: (h3) the revision reported was likely the result of an insufficient bond between the implant and the bone, patient¿s bone quality, or a combination of the two, which led to aseptic (non-infected) loosening of the glenoid baseplate and instability.However, this cannot be confirmed as the devices were not available for evaluation.(h4) device manufacture date: 08-feb-2019.
 
Additional Manufacturer Narrative
Section h10: (g5) pma/510(k)number: k063569.Section h11: *the following sections have corrected information: (d4) catalog number: 320-15-01, serial number: (b)(6), expiration date: 05-feb-2029, unique identifier (udi) #: (b)(4).(d11) concomitant device(s): ** added glenosphere** 36mm glenosphere (cat# 320-01-36 / sn#: b)(6)).Reverse tray (cat# 320-10-00 / sn#: b)(6)).Humeral liner (cat# 320-36-13 / sn#: b)(6)).Torque defining screw kit (cat# 320-20-00 / sn#: b)(6)).Reverse tray adapter plate (cat# 320-10-10 / sn#: b)(6)).Torgue defining screw kit (cat# 320-20-00 / sn#: b)(6)).Rev locking screw (cat# 320-15-05 / sn#: b)(6)).Compress lck cap (cat# 320-20-18 / sn#: b)(6)).Compress lck cap (cat# 320-20-26 / sn#: b)(6)).Compress lck cap (cat# 320-20-22 / sn#: b)(6)).Compress lck cap (cat# 320-20-18 / sn#: b)(6)).Compress lck cap (cat# 320-20-30 / sn#: b)(6)).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
EQUINOXE
Common Device Name
36MM GLEN TRL
Manufacturer (Section D)
EXACTECH, INC.
2320 nw 66 court
gainesville FL 32653
MDR Report Key9134811
Report Number1038671-2019-00474
Device Sequence Number3587256
Product Code KWT
UDI-Device Identifier10885862086372
UDI-Public10885862086372
Combination Product (Y/N)N
Initial Reporter CountryAS
PMA/510(K) Number
K063569
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2019
Device Explanted Year2019
Reporter Type Manufacturer
Report Source company representative,health
Initial Reporter Occupation Physician
Type of Report Initial,Followup,Followup
Report Date (Section B) 03/31/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Model Number320-01-36
Device Catalogue Number320-15-01
Was Device Available for Evaluation? No
Initial Date Received by Manufacturer 09/11/2019
Supplement Date Received by Manufacturer02/06/2020
03/25/2020
Initial Report FDA Received Date09/30/2019
Supplement Report FDA Received Date02/06/2020
03/31/2020
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Hospitalization; Required Intervention;
-
-