|
A manufacturing record evaluation was performed for the finished device mm6944, and no non-conformances were identified.Additional summary: received eight unopened samples of product code sxpp1a419, lot mm6944 for evaluation.The complaint sample was not returned.During the visual inspection of samples, no defects were found on the packages.The samples were opened, and the swage and attachment area were noted to be as expected.No needles breakage was observed.The sutures were dispensed without problems and examined along of the strand and no defects were observed.Also, the samples were tested by resistance test to needle and met the requirements.Per the conditions of the sample received, no performance breakage needle were found, representative samples returned to support investigation of 4 quantity of device issues captured in reports 2210968-2019-88228, 2210968-2019-88230, 2210968-2019-88231 and 2210968-2019-88232.Analysis results have been included in follow-up reports for each.
|