| Model Number |
434890 |
| Medical Device Problem Code |
Device Dislodged or Dislocated (2923)
|
| Health Effect - Clinical Code |
No Consequences Or Impact To Patient (2199)
|
| Date of Event |
09/05/2019
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
The pk papyrus 4.0/15 was selected for the treatment of a perforation in the lad.As it was not possible to place the pk papyrus proximal to previously implanted stents it was decided to remove the pk papyrus.During withdrawal the stent dislodged from the balloon inside the teleflex guideliner extension catheter.
|
| |
|
Additional Manufacturer Narrative
|
|
Neither the complaint instrument nor the angiographic material was provided for analysis.Therefore no technical investigation on the subject could be performed.The production documentation was reviewed to establish whether a deviation from the manufacturing process could be the cause for the reported event.Review of the production documentation confirmed that the instrument was manufactured according to specifications and passed all in-process and final inspections.During final inspection, every stent system undergoes visual inspection to ensure correct stent embedding and homogeneous crimping of the stent.Further, the stent outer diameter is verified to 100 percent and the stent retention force of a defined amount of samples is tested from every lot.The device in question was manufactured according to specifications and successfully passed all in-process controls as well as the final inspection.Based on the conducted investigations of the device being subject to this complaint, no material or manufacturing related root cause was determined.
|
| |
|
Search Alerts/Recalls
|