Catalog Number 242400
Medical Device Problem Code Intermittent Continuity (1121)
Health Effect - Clinical Code Not Applicable (3189)
Date of Event 08/30/2019
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Product complaint # (b)(4).Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.(b)(6).Udi: (b)(4).The lot number is unknown.
 
Event or Problem Description
It was reported by the affiliate via email that during a percutaneous nephrolithotomy (pcnl) procedure the tck1 hd camera head stopped relaying the image or video on to the attached monitor and displayed.The surgeon managed to complete the surgery by using another alternative lap set available with the hospital.Additional information provided by the affiliate reported the two devices were involved in two separate incidents.The affiliate later reported the alleged malfunction was found during a pcnl procedure, this particular camera stopped working the case was completed because hospital is equipped with another laparoscopy set.The surgical team was able to replace the malfunctioning camera an alternative set.It was also reported the alleged malfunction caused a 10 to 15 minute surgical delay which did not cause any patient or user harm.
 
Additional Manufacturer Narrative
Product complaint # (b)(4).Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.Investigation summary: the complaint device was received at the service center and evaluated.During evaluation, no defect was found, therefore this complaint cannot be confirmed.As the reported problem was not confirmed, a root cause for the issue that was experienced by the user cannot be determined.No nonconformance's were identified for this part-serial number combination per qlik query executed on 11/21/2019.At this point in time, no corrective action is required, and no further action is warranted.However, depuy synthes mitek will continue to track any related complaints within this device family as a means of monitoring the extent with which this complaint is observed in the field.Udi: (b)(4).
 
Additional Manufacturer Narrative
Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.H3, h6: the device was received, the investigation is in progress, no conclusion could be drawn at the time of filing this report.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TCK1 HD CAMERA HEAD
Common Device Name
ENDOSCOPIC VIDEO CAMERA
Manufacturer (Section D)
MEDOS INTERNATIONAL SàRL
chemin-blanc 38
le locle 02400
SZ  02400
MDR Report Key9136187
Report Number1221934-2019-58682
Device Sequence Number16940362
Product Code FWF
UDI-Device Identifier10886705026258
UDI-Public10886705026258
Combination Product (Y/N)N
Initial Reporter CountryIN
PMA/510(K) Number
EXEMPT
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source foreign,other
Initial Reporter Occupation Other
Type of Report Initial,Followup,Followup
Report Date (Section B) 08/31/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Catalogue Number242400
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer09/30/2019
Initial Date Received by Manufacturer 08/31/2019
Supplement Date Received by Manufacturer11/21/2019
11/26/2019
Initial Report FDA Received Date09/30/2019
Supplement Report FDA Received Date11/21/2019
11/27/2019
Patient Sequence Number1
-
-