Manufacturer Narrative
|
Product complaint # (b)(4).Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.(b)(6).Udi: (b)(4).The lot number is unknown.
|
|
Event Description
|
It was reported by the affiliate via email that during a percutaneous nephrolithotomy (pcnl) procedure the tck1 hd camera head stopped relaying the image or video on to the attached monitor and displayed.The surgeon managed to complete the surgery by using another alternative lap set available with the hospital.Additional information provided by the affiliate reported the two devices were involved in two separate incidents.The affiliate later reported the alleged malfunction was found during a pcnl procedure, this particular camera stopped working the case was completed because hospital is equipped with another laparoscopy set.The surgical team was able to replace the malfunctioning camera an alternative set.It was also reported the alleged malfunction caused a 10 to 15 minute surgical delay which did not cause any patient or user harm.
|
|
Manufacturer Narrative
|
Product complaint # (b)(4).Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.Investigation summary: the complaint device was received at the service center and evaluated.During evaluation, no defect was found, therefore this complaint cannot be confirmed.As the reported problem was not confirmed, a root cause for the issue that was experienced by the user cannot be determined.No nonconformance's were identified for this part-serial number combination per qlik query executed on 11/21/2019.At this point in time, no corrective action is required, and no further action is warranted.However, depuy synthes mitek will continue to track any related complaints within this device family as a means of monitoring the extent with which this complaint is observed in the field.Udi: (b)(4).
|
|
Manufacturer Narrative
|
Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.H3, h6: the device was received, the investigation is in progress, no conclusion could be drawn at the time of filing this report.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
|
|