• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: TISSUE SCIENCE LABORATORIES PERMACOL; MESH, SURGICAL

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

TISSUE SCIENCE LABORATORIES PERMACOL; MESH, SURGICAL Back to Search Results
Model Number P101010
Medical Device Problem Codes Loss of or Failure to Bond (1068); Material Integrity Problem (2978); Adverse Event Without Identified Device or Use Problem (2993)
Health Effect - Clinical Codes Adhesion(s) (1695); Injury (2348)
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
If information is provided in the future, a supplemental report will be issued.
 
Event or Problem Description
The patient¿s attorney alleged a deficiency against the device.The product was used for therapeutic laparoscopic treatment of a hernia.It was reported that after implant, the patient experienced mesh contracted and pulled away from the fascial edge, extensive adhesions, and small serosal tear/enterotomy repaired with suture.Post-operative patient treatment included revision surgery.The device had been used with endotak.
 
Additional Manufacturer Narrative
Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PERMACOL
Common Device Name
MESH, SURGICAL
Manufacturer (Section D)
TISSUE SCIENCE LABORATORIES
victoria house, victoria road
aldershot, hampshire GU11 1EJ
GB  GU11 1EJ
MDR Report Key9136740
Report Number9617613-2019-00162
Device Sequence Number16817906
Product Code FTM
UDI-Device Identifier10884523000108
UDI-Public10884523000108
Combination Product (Y/N)N
Initial Reporter StateNY
Initial Reporter CountryUS
PMA/510(K) Number
K120605
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2007
Reporter Type Manufacturer
Report Source consumer
Initial Reporter Occupation Non-Healthcare Professional
Type of Report Initial,Followup
Report Date (Section B) 10/04/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date06/14/2009
Device Model NumberP101010
Device Catalogue NumberP101010
Device Lot Number06B11-9
Was Device Available for Evaluation? No
Initial Date Received by Manufacturer 09/04/2019
Supplement Date Received by Manufacturer09/04/2019
Initial Report FDA Received Date09/30/2019
Supplement Report FDA Received Date10/04/2019
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
ENDOTAK
Outcome Attributed to Adverse Event Required Intervention;
Patient Age47 YR
Patient Weight137
-
-