| Model Number |
MT22719-BLU |
| Medical Device Problem Code |
Inappropriate or Unexpected Reset (2959)
|
| Health Effect - Clinical Code |
No Consequences Or Impact To Patient (2199)
|
| Date of Event |
08/22/2019
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
(b)(4).
|
| |
|
Event or Problem Description
|
|
It was reported that the receiver initialized without a manual restart.Product was received but is pending evaluation.A follow up report will be submitted upon completion.No injury or medical intervention was reported.
|
| |
|
Additional Manufacturer Narrative
|
|
(b)(4).
|
| |
|
Event or Problem Description
|
|
It was reported that the receiver initialized without a manual restart.The product was evaluated.An external visual inspection was performed and passed.A receiver charge and boot was performed and failed due to receiver perpetually rebooting.A receiver download was performed and no data was provided for evaluation.The receiver case was opened, and an internal visual inspection was performed and passed.Confirmation of the allegation and probable cause could not be determined.No injury or medical intervention was reported.
|
| |
|