• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: SMITH & NEPHEW, INC. LEGION PS NP FEM SZ 6 RT; PROSTH, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONST, CEMENTED, POLY/METAL/POLY

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SMITH & NEPHEW, INC. LEGION PS NP FEM SZ 6 RT; PROSTH, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONST, CEMENTED, POLY/METAL/POLY Back to Search Results
Model Number 71423236
Medical Device Problem Code Unstable (1667)
Health Effect - Clinical Code Injury (2348)
Date of Event 09/03/2019
Type of Reportable Event Serious Injury
Event or Problem Description
It was reported that a revision surgery was reported due to instability and pain.The femoral component, the baseplate and the insert were replaced.
 
Additional Manufacturer Narrative
It was reported that a revision surgery was reported due to instability and pain.The associated genesis ii ps high flexion insert, genesis ii cemented tibial base plate and legion ps femoral component, used in treatment, were not returned.Thus, no product analysis could be performed.Our investigation including a review of the manufacturing records for the listed batches did not reveal any deviation from the standard manufacturing processes.A review of the complaint history for the listed parts revealed no prior complaints for the listed failure mode with the same batch number.At this time, we have no reason to suspect that the products failed to meet any product specifications at the time of manufacture.A relationship, if any, between the devices and the reported incident could not be corroborated.No medical documents were received for investigation.Therefore no medical assessment can be performed at this time.A review of instructions for use revealed the alleged failure is previously identified in the potential adverse events.Per our risk assessment, possible causes could include but not limited to device alignment, or joint laxity.Based on this investigation, the need for corrective action is not indicated.Without the return of the actual products involved and no patient records available, our investigation of this report is inconclusive.No further investigation is warranted for this complaint; however, we will continue to monitor for future complaints and investigate as needed.Should additional information be received, this complaint will be reopened and reevaluated.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
LEGION PS NP FEM SZ 6 RT
Common Device Name
PROSTH, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONST, CEMENTED, POLY/METAL/POLY
Manufacturer (Section D)
SMITH & NEPHEW, INC.
1450 brooks road
memphis TN 38116
MDR Report Key9137591
Report Number1020279-2019-03515
Device Sequence Number19528599
Product Code JWH
UDI-Device Identifier00885556031001
UDI-Public00885556031001
Combination Product (Y/N)N
Initial Reporter CountryAS
PMA/510(K) Number
K060742
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,foreig
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 02/01/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date11/05/2021
Device Model Number71423236
Device Catalogue Number71423236
Device Lot Number11LM03999
Was Device Available for Evaluation? No
Initial Date Received by Manufacturer 09/04/2019
Supplement Date Received by Manufacturer01/28/2020
Initial Report FDA Received Date09/30/2019
Supplement Report FDA Received Date02/01/2020
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Hospitalization; Required Intervention;
-
-