| Catalog Number |
49401020 |
| Medical Device Problem Code |
Loose or Intermittent Connection (1371)
|
| Health Effect - Clinical Codes |
No Consequences Or Impact To Patient (2199); No Known Impact Or Consequence To Patient (2692)
|
| Date of Event |
09/03/2019
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Upon completion of investigation, additional information will be provided in a supplemental report.
|
| |
|
Event or Problem Description
|
|
As reported: "during the operation, the needle clamp of hoffman ii product was not tightly clamped.After the nail is nailed, when connecting the pin rod chuck, it is found that the pin chuck is not clamped tightly and falls off after screwing.Subsequently, the operation was performed by other batch products." procedure was completed successfully with 15 minutes surgical delay and other batch of product was used to complete the surgery with no adverse consequences reported.
|
| |
|
Event or Problem Description
|
|
As reported: "during the operation, the needle clamp of hoffman ii product [.] was not tightly clamped.After the nail is nailed, when connecting the pin rod chuck, it is found that the pin chuck is not clamped tightly and falls off after screwing.Subsequently, the operation was performed by other batch products." procedure was completed successfully with 15 minutes surgical delay and other batch of product was used to complete the surgery with no adverse consequences reported.
|
| |
|
Additional Manufacturer Narrative
|
|
The reported event could be confirmed.Inspection of the returned device showed that the pin to rod coupling is wrongly assembled.Based on investigation, the root cause was attributed to be manufacturing related.The failure was caused by a wrong assembly of the part.An nonconformity was raised to address the issue.If any further information is provided, the investigation report will be updated.
|
| |
|