| Model Number |
MMT-751NAL |
| Medical Device Problem Code |
Device Difficult to Program or Calibrate (1496)
|
| Health Effect - Clinical Code |
No Consequences Or Impact To Patient (2199)
|
| Date of Event |
09/25/2019
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
(b)(4).Currently it is unknown whether or not the device may have caused or contributed to the event as no product has been returned.No conclusion can be drawn at this time.We therefore consider this report complete to the best of our knowledge.
|
| |
|
Event or Problem Description
|
|
The customer reported via phone call that the insulin pump bolus button sticking.The customer's blood glucose value was unknown.The insulin pump will not be returned for analysis.
|
| |
|
Additional Manufacturer Narrative
|
|
No button error alarm noted during testing.Device had intermittent buttons response due to flattened (creased) act button dome switch.No unlocked lcd keypad connector noted.
|
| |
|
Search Alerts/Recalls
|