| Model Number |
MMT-722NAP |
| Medical Device Problem Codes |
Mechanical Problem (1384); Priming Problem (4040)
|
| Health Effect - Clinical Code |
No Consequences Or Impact To Patient (2199)
|
| Date of Event |
12/21/2018
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
(b)(4).Currently it is unknown whether or not the device may have caused or contributed to the event as no product has been returned.The device will be returned for analysis and further information will follow once the analysis has been completed.No conclusion can be drawn at this time.
|
| |
|
Event or Problem Description
|
|
The customer reported via phone call that the insulin pump had prime anomaly and motor error alarm.The customer¿s blood glucose level was unknown.The customer also reported that the insulin was squirting during manual prime process.The customer stated that the drive support cap was normal.Customer was able to complete insulin pump rewind.Customer stated its shooting out the insulin.The customer was advised that the insulin pump needed to be replaced and was advised to discontinue use of the insulin pump and to revert to the backup plan.The insulin pump will be returned for analysis.
|
| |
|
Search Alerts/Recalls
|